Catheter, Cannula, and Guidewire Measurement: Choosing the Right Laser Micrometer Setup

Updated August 21, 2026

A laser micrometer does not measure a catheter, cannula, or guidewire in the abstract. It measures the silhouette that the part presents inside the measurement field. If the part sags, wanders, twists, tilts, or is squeezed by the fixture, a capable sensor can report a repeatable number that does not represent the intended dimension. The best setup therefore combines the right LaserLinc micrometer with the right support, motion, inspection sequence, software, and validation method.

Application focus: This article begins with the part and measurement decision, not a model number. LaserLinc combines laser micrometers with fixtures, guided part handling, ultrasonics, surface inspection, automation, and software for complete measurement systems. The company combines Axion and Triton laser gauges with purpose-built fixtures, roller and feed-through guides, BenchLinc systems, Metron automation, UltraGauge ultrasonics, FlawSense surface inspection, and Total Vu software. Gauge Advisor helps translate the drawing, part behavior, workflow, and quality requirement into the most practical LaserLinc architecture.

Connected process path: Start with the required feature and part presentation, then combine the sensor, fixture, motion, and software needed for repeatable catheter, cannula, or guidewire inspection.

Related next steps: medical-device measurement and inspection, medical-tubing measurement selector, laser micrometers, medical-tubing wall-thickness measurement, and medical-tubing surface-defect detection.

Interactive tool unavailable? The complete article and application guidance remain available below.
LaserLinc laser micrometer measuring tubing supported by adjustable roller guides
The sensor is only one part of the measurement. Product support, centering, orientation, motion, and the inspection recipe determine whether the displayed OD and ovality represent the intended location and part condition.
The complete measurement system

Reliable catheter measurement has five connected parts

Teams often begin by asking whether they need one, two, or three laser axes. That matters, but it is not the first decision. Start by defining the feature being measured and the part state, then build the complete measurement chain around it.

1Part and feature being measuredOD, ovality, width, taper, transition, wall, concentricity, surface defect, or a full length-positioned profile.
2Support and datumV-block, feed-through guide, roller guide, chuck, cone, centering cup, mandrel, or application-specific holder.
3Measurement technologyAxion or Triton for outside geometry, UltraGauge for wall, FlawSense for surface topography, or a combined architecture.
4Motion and positionManual feed, controlled rotation, encoded translation, automated full-length scan, or continuous inline movement.
5Decision and recordsTotal Vu recipes, statistics, profile zones, pass or fail logic, reports, integration, and retained evidence.

LaserLinc describes its laser micrometers as adaptable components within integrated measurement solutions, and its sample-inspection platforms add part holding, motion, automated recipes, and Total Vu reporting rather than leaving the user to assemble a loose collection of instruments.[1][4] That distinction is important in medical manufacturing, where the fixture and inspection sequence can contribute as much variation as the sensor itself.

Define the dimension before choosing the fixture. A laser micrometer directly measures outside silhouette dimensions. It does not directly see ID, wall thickness, concentricity, or a buried feature. Those requirements may need BenchLinc OD/ID, BenchLinc UT, inline UltraGauge, or another complementary method.
Interactive application review

How should the part be presented to the measurement system?

Select the product that most closely matches the application. The result identifies the likely support method, measurement architecture, and the first condition I would verify with representative samples.

Soft catheter or microtube: stabilize without reshaping it

A small flexible tube may sag below the measurement centerline, wander laterally, or flatten when it contacts a fixture. A feed-through guide, carefully positioned rollers, or a purpose-built through-hole support can improve repeatability, but the guide clearance and contact condition must be proven on the actual material.

  • Start with the lowest-contact support that keeps the measurement location stable.
  • Check whether friction, tack, curve memory, temperature, or guide contact changes the displayed diameter.
  • Use a short sample study with the fixture installed, not only a free-air gauge repeatability test.
Feed-through or rollersSample testingLow contact force
Why fixturing changes the result

Five ways part presentation can create measurement error

NIST dimensional-metrology guidance identifies alignment, artifact geometry, elastic deformation, support geometry, temperature, and motion as recurring contributors to measurement uncertainty.[11][12] On a catheter or guidewire, those effects often appear as the following practical problems.

Sag

The measurement location drops below the intended centerline. A long unsupported span can also bend differently from sample to sample.

Lateral wander

The part moves through different regions of the field or contacts guides unpredictably, producing a noisy or biased trend.

Angular misalignment

The part axis is not normal to the measurement plane. The projected silhouette can become larger than the true cross section.

Contact deformation

A soft tube can flatten under a clamp, guide, roller, chuck, drive band, or hand pressure. Repeatable force is not the same as negligible force.

Orientation bias

An oval or asymmetric part can produce a different result when placed in a different rotational orientation. The fixture may lock in that bias.

A simple free-span test is useful before ordering a fixture: place the part at the intended measurement height, vary the unsupported length, and observe whether the reading changes as the part is released, rotated, or lightly moved. That test can reveal whether support is needed, but it does not establish the final fixture, uncertainty, or measurement capability.

The practical goal is not maximum restraint. It is enough restraint to make the part location and orientation repeatable without altering the dimension being measured. The correct balance changes with durometer, wall thickness, diameter, curvature, braid, coating, temperature, and total unsupported length.
LaserLinc fixturing options

Choose the support architecture from part behavior and inspection motion

LaserLinc publishes a broad range of V-block, feed-through, roller, chuck, encoder, Metron, and application-specific holding options. The important choice is not which accessory looks most familiar. It is which support creates the most repeatable part state for the required inspection.

Stationary cut part

V-block fixture

A V-shaped groove self-centers many round, reasonably stiff parts and gives the operator a repeatable placement surface. LaserLinc offers standard and long V-block arrangements for cut parts.[5]

  • Strong fit for rigid or semi-rigid round samples
  • Simple loading and visual access
  • Can support repeatable point measurements
Watch: oval parts can settle in a preferred orientation, and very soft tubing can conform to the groove.
Small flexible tubing

Feed-through guide

LaserLinc feed-through guides mount to a V-block and use replaceable guide sizes to center and support small flexible tubing through the measurement field.[5]

  • Reduces sag and lateral movement
  • Supports manual feed-through inspection
  • Useful where an open V-block does not hold position
Watch: clearance, friction, contamination, curve memory, and tack can influence the part.
Moving or continuously fed part

Roller or throat guide

Throat-mount, body-mount, plate-mount, rail-mount, and adjustable roller guides support tubing or wire while allowing movement through the micrometer.[5]

  • Good for translation through the field
  • Adjustable alignment for size changes
  • Can be paired with encoders and automation
Watch: roller runout, product tension, guide spacing, debris, and contact angle can become measurement variables.
Automated rotation

Spring-loaded chuck or part rotation

BenchLinc OD uses a spring-loaded chuck and automatic rotation for repeatable outside-geometry profiles. Metron can add motorized part rotation for circumferential inspection.[4][6]

  • Reduces manual orientation differences
  • Supports multiple angular measurements
  • Useful for ovality and asymmetric features
Watch: chuck force, eccentric clamping, runout, end condition, and torsional windup must be validated.
Long or multi-zone part

Metron part holders

Metron options include centering cones, magnet V-blocks, centering cups, toggle-clamp V-blocks, cone holders, zero-chucks, throat guides, and motorized part holders.[6]

  • Supports vertical or horizontal full-length scanning
  • Accommodates hubs, mandrels, varying ends, and long parts
  • Can combine translation and rotation
Watch: applied tension, vertical weight, clamp location, part straightness, and transition definitions.
Internal geometry

Mandrel and ultrasonic fixtures

When ID, wall, or concentricity controls the requirement, the support method may be part of a BenchLinc OD/ID or BenchLinc UT measurement architecture rather than a simple laser fixture.[4]

  • BenchLinc OD/ID for short polymer samples
  • BenchLinc UT for rigid and metallic tubing profiles
  • Inline UltraGauge for continuous extrusion
Watch: do not report a calculated or datum-based result as though the laser directly measured the inner surface.
LaserLinc Triton 330 three-axis laser micrometer
Triton: three-axis outside-geometry measurement for round and elliptical parts, with fixture and guide options selected for the application.
LaserLinc vertical Metron system for full-length catheter and guidewire inspection
Metron: automated full-length inspection with part holding, translation, optional rotation, profile recipes, statistics, and reports.
LaserLinc BenchLinc UT for rigid cannula hypotube and medical tubing measurement
BenchLinc UT: an integrated ultrasonic and optional laser system for wall, concentricity, OD, ovality, and calculated ID on compatible rigid tubing.
Interactive planning tool

Build a preliminary LaserLinc measurement and fixturing path

This selector identifies a practical starting architecture from the part behavior, inspection length, feature being measured, and desired automation. Final selection still requires dimensions, tolerances, photographs or drawings, representative samples, and a review of the actual workflow.

More than a micrometer company

LaserLinc can build the complete inspection architecture

A common purchasing mistake is to compare only micrometer resolution, rate, and range. Those specifications matter, but they do not determine whether a medical-device team receives a reliable workflow. LaserLinc positions its systems as adaptable, integrated measurement, automation, and process solutions rather than isolated sensors.[7][8]

1. SensorFit-for-purpose measurement technology

Axion for one- or two-axis applications, Triton for three-axis round and elliptical measurement, UltraGauge for wall, and FlawSense for surface topography.

2. FixturePart holding around the real product

V-blocks, feed-through guides, rollers, chucks, mandrels, cones, centering cups, throat guides, and custom holders selected from part behavior.

3. MotionTranslation, rotation, and position

Manual feed, encoder feedback, automatic rotation, full-length Metron scanning, inline motion, markers, cutters, and application-specific automation.

4. SoftwareTotal Vu measurement workflow

Recipes, displays, calculated values, profile zones, SPC, pass or fail decisions, reports, data export, communications, and control tools.

5. VerificationSample testing and method development

Virtual demonstrations with customer parts, application engineering, correlation planning, calibration support, and configuration review before release.

6. LifecycleWarranty and live support

LaserLinc publishes a typical four-year warranty for laser micrometers and accessories, no-charge phone and email support for the original purchaser, remote Quick Support, and long-term software support under the stated terms.[9]

This is where custom solutions create value. The goal is not custom hardware for its own sake. It is a purpose-fit combination of proven sensors, fixtures, motion components, software tools, and support that removes operator improvisation from a critical inspection.
Fit-for-purpose comparison

Match the part and feature being measured to the LaserLinc system

Fixturing cannot make the wrong measurement technology appropriate. Use this table to separate simple OD checks from full-length profiling, wall measurement, and surface inspection.

Starting architectureBest-fit part and workflowTypical support or holdingWhat it measures and important boundary
Point or moving ODCatheter shafts, tubing, guidewires, cannulas, mandrels, and other parts needing outside dimensions at one or several positions.V-block, feed-through guide, throat or roller guide, adjustable stand, or application-specific holder.Direct outside silhouette dimensions and ovality estimates according to axis coverage. Does not directly measure wall, ID, concentricity, or hidden defects.
Repeatable bench ODRecurring QA checks on short or manageable samples where a standardized recipe and automated rotation improve consistency.V-block or through-hole guide for BenchLinc V; spring-loaded chuck and automatic rotation for BenchLinc OD.OD, ovality, angular profile, statistics, tolerances, and reports. Internal geometry requires another architecture.
Full-length profileLong catheters, guidewires, tapers, proximal and distal zones, transition locations, and complex finished-part profiles.Vertical or horizontal configuration with cones, V-blocks, cups, chucks, throat guides, motorized part holder, and optional rotation.Length-positioned outside geometry and configured features; can integrate other LaserLinc technologies. Part holding and profile definitions must match the drawing.
Short polymer samplesShort cut polymer tubing samples with an accessible lumen and a suitable mandrel.Special mandrel, controlled rotation, motor bands, and integrated load measurement.Laser OD and ovality, wall relative to the mandrel datum, and calculated ID. Soft-part deformation and mandrel fit require validation.
Rigid wall profileCannulas, hypotubes, NiTi, stainless steel, and compatible rigid tubing needing wall and concentricity along length.Purpose-built ultrasonic fixture and part guides; optional Triton for OD and ovality.Direct ultrasonic wall and concentricity, with optional outside geometry and calculated ID. Acoustic feasibility and sample handling must be confirmed.
Surface topographyCatheters, guidewires, coated shafts, tubing, and components where pits, scratches, cracks, bumps, or other raised and recessed features control acceptance.Inline or offline guiding selected to maintain centering without masking the defect surface.Three-dimensional visible surface geometry and outside dimensions. Appearance-only, internal, coating-chemistry, or functional defects may need another method.

The Medical Tubing Measurement Selector provides a broader starting path when the project includes several features being measured or both inline and offline inspection.

From application to supported system

Gauge Advisor helps teams choose the approach, not just the gauge

A useful application review should reduce risk before a quotation is issued. That means understanding the product drawing, how the part behaves, which values are direct or calculated, where the inspection belongs, how operators will load the part, and what evidence the quality team needs to retain.

Matthew’s role in the project

  • Translate the product and drawing into a defined feature being measured and inspection sequence.
  • Compare V-block, feed-through, roller, BenchLinc, Metron, ultrasonic, and defect-inspection paths.
  • Identify where sample testing or custom holding is needed.
  • Coordinate LaserLinc application engineering, demonstrations, and feasibility work.
  • Develop the quotation and review integration, training, and support expectations.
  • Remain involved after the sale for applications questions and local coordination.

What I need from the customer

The fastest route to a good recommendation is a small, complete application package rather than a request for a generic micrometer quote.

  • Part drawings or dimensioned sketches with measurement locations
  • Minimum, nominal, and maximum OD, wall, ID, and tolerances
  • Part length, stiffness, durometer, curvature, coating, braid, and end features
  • Current inspection method, fixture, measurement problems, and MSA data
  • Desired cycle time, operators, recipes, reporting, and integration
  • Representative good, borderline, and known-difficult samples

Practical outcome: the recommendation may be a simple fixture and micrometer, an automated BenchLinc or Metron system, a combined ultrasonic and laser system, or a decision to test the part before specifying anything.

Validate the complete method

Fixture repeatability belongs in the measurement-system study

Calibration of the laser scale does not validate how the customer part is supported, loaded, translated, rotated, or interpreted. The complete method should include the actual fixture, part family, recipe, operators, environment, and acceptance decision.

1Define the feature being measured

State the exact location, orientation, section, profile zone, formula, and whether the reported value is direct, derived, averaged, or inferred.

2Control the part state

Document temperature, conditioning, cleaning, support span, curve memory, guide clearance, clamp force, tension, rotation, and loading method.

3Verify across the range

Use suitable reference standards for the micrometer and representative parts spanning size, stiffness, color, surface, tolerance, and difficult conditions.

4Study the complete workflow

Evaluate repeatability, reproducibility, bias, stability, fixture changes, recipe selection, handling, reporting, and borderline decisions.

NIST defines metrological traceability as a property of a measurement result established through a documented unbroken chain of calibrations, each contributing to measurement uncertainty.[13] It is not an automatic label attached to a micrometer. For medical-device manufacturers, FDA’s Quality Management System Regulation became effective February 2, 2026 and incorporates ISO 13485:2016 by reference.[14] LaserLinc equipment can support a controlled quality process, but the manufacturer remains responsible for validating the intended use, procedures, records, maintenance, software configuration, and changes.

Do not validate only the easiest part. Include the softest catheter, longest guidewire, smallest and largest sizes, tightest tolerance, highest curvature, most reflective or transparent material, difficult end condition, and any product known to behave differently in the fixture.
Avoid false confidence

Common catheter and guidewire measurement setup mistakes

Buying the micrometer before defining the support method

The part can fit inside the measurement range and still be impossible to present repeatably. Review stiffness, length, curve memory, contact sensitivity, access, and motion before finalizing the gauge and stand.

Using a V-block for every round product

A V-block is simple and effective for many rigid or semi-rigid parts. A very soft tube may conform to the groove, while an oval part may settle in a preferred orientation and hide rotational variation.

Assuming a feed-through guide is noncontact

The laser measurement is noncontact, but the guide is not. Guide diameter, length, material, friction, cleanliness, curve memory, coating, and manual feed force can influence the part.

Calling three-axis diameter a full surface scan

Triton provides three outside-diameter views and improved angular coverage for round and elliptical products. It does not create a complete three-dimensional surface map. Use FlawSense when the defect can occupy a small local region.

Using OD as proof of wall, ID, or concentricity

A tube can maintain outside diameter while its lumen moves or its wall changes. Use the appropriate mandrel-based or ultrasonic system when internal geometry controls acceptance.

Applying too much tension to straighten a guidewire

Tension can improve presentation while also stretching, aligning, or reshaping the part. Define and control the applied condition, and verify that it does not change the dimension or profile being released.

Clamping on a hub, taper, coating, or functional region

A convenient holding point can damage the product, change its shape, or move the measurement datum. The fixture should protect critical surfaces and preserve access to every required zone.

Validating the sensor but not the fixture recipe

Calibration confirms the measurement scale under defined conditions. It does not prove operator loading, guide setup, chuck force, scan speed, profile-zone definitions, or report logic.

Expecting a standard accessory to fit every future part

A fixture should cover the intended product family without becoming so adjustable that setup is ambiguous. Define change parts, recipes, verification checks, and when a new holder or sample study is required.

Prepare the application review

Catheter, cannula, and guidewire measurement checklist

  • Product type and constructionCatheter, tubing, cannula, hypotube, guidewire, mandrel, shaft, braid, layers, coating, and material.
  • Complete dimensional rangeMinimum, nominal, and maximum OD, width, height, wall, ID, and feature dimensions.
  • Tolerances and formulasUpper and lower limits, ovality definition, taper or transition rules, profile zones, and minimum-wall criteria.
  • Length and part behaviorTotal length, supported and unsupported span, stiffness, durometer, sag, curvature, memory, and straightness.
  • Surface and contact sensitivityTack, lubricity, coatings, markers, hydrophilic surfaces, braid texture, scratches, and areas that cannot be touched.
  • Current methodExisting gauge, fixture, operator steps, reference standard, sample preparation, cycle time, and known disagreement.
  • Inspection workflowPoint check, feed-through, rotation, full-length scan, inline measurement, batch size, operators, and desired automation.
  • Data requirementsPass or fail, statistics, profiles, reports, CSV, OPC UA, PLC, MES, audit trail, permissions, and record retention.
  • Representative samplesGood, borderline, reject, softest, longest, smallest, largest, most curved, and most difficult end condition.
  • Commercial and support scopeTiming, quotation needs, virtual demo, sample testing, installation, training, validation support, and service expectations.
Continue the measurement plan

Related Gauge Advisor medical measurement resources

Frequently asked questions

Catheter, cannula, and guidewire fixturing FAQs

Do I always need a fixture with a laser micrometer?

No. A short, straight, sufficiently rigid part may be positioned repeatably without a dedicated fixture. Add support when sag, wander, tilt, rotation, vibration, operator handling, or cycle time prevents a capable measurement method. Prove the need with the actual part rather than assuming either free-air or heavy restraint is always better.

When is a V-block the best choice?

A V-block is a strong starting point for round, cut-to-length parts that are stiff enough to rest in the groove without changing shape. It provides simple self-centering and repeatable placement. Validate orientation effects for oval parts and deformation for soft tubing.

When should I use a feed-through guide?

Evaluate a feed-through guide when small flexible tubing cannot remain centered or steady during manual translation. The guide should provide enough clearance for smooth movement while controlling position. Its contact, friction, cleanliness, and effect on the part must be included in the method study.

Are roller guides only for inline extrusion?

No. Roller and throat guides can support inline product, manual feed-through inspection, encoded translation, and automated sample systems. Guide type, spacing, adjustment, runout, product tension, and contact condition determine whether the arrangement is suitable.

Should I choose a dual- or triple-axis laser micrometer?

Choose axis count from shape, angular coverage, size range, installation space, rate, and the consequence of missing orientation-dependent variation. Triton three-axis systems are well suited to round and elliptical products, while Axion one- and two-axis systems can be better for non-round, constrained, hot, multi-strand, or value-focused applications.

How do I inspect a long catheter or guidewire?

Use Metron when the requirement is a length-positioned profile across proximal, body, taper, transition, distal, or other defined zones. Vertical or horizontal configuration, part holders, optional rotation, scan length, profile definitions, and reporting are selected from the actual product.

Can a laser micrometer measure cannula wall thickness?

Not directly. A laser micrometer measures outside geometry. BenchLinc UT is the stronger starting path for compatible rigid cannulas and hypotubes when direct ultrasonic wall and concentricity are required, with an optional Triton for OD, ovality, and calculated ID.

Can LaserLinc build a custom fixture?

LaserLinc offers a broad set of standard accessories and integrated part-holding options, and it supports adaptable, application-specific systems. The right approach may use standard components, configured combinations, or custom-engineered elements under the final project scope.

What is Gauge Advisor’s role?

Gauge Advisor is the authorized LaserLinc sales and applications support partner. Matthew helps define the measurement problem, compare systems and fixtures, coordinate sample testing and LaserLinc engineering, prepare quotations, review integration, and support the application after purchase.

What samples should I send?

Send representative good parts plus the smallest, largest, softest, longest, most curved, most reflective or transparent, tightest-tolerance, borderline, and known-difficult samples. Include drawings, locations, current measurements, and any fixture-related disagreement.

Technical basis

References and source notes

The references remain collapsible so this focused commercial article is easy to scan. They open automatically when printing.

Open technical references and source notes16 sources

Gauge Advisor is the authorized LaserLinc sales and applications support partner. LaserLinc sources are used for current micrometer, fixture, sample-inspection, software, warranty, and integrated-solution information. NIST, FDA, and ISO sources provide noncompetitive metrology, quality-system, catheter, guidewire, and cannula context. No competitive measurement-equipment manufacturer is cited.

  1. LaserLinc, Laser Micrometers. Current description of outside-dimension measurement, Axion and Triton categories, noncontact operation, and integrated measurement solutions.
  2. LaserLinc, Triton Three-Axis Laser Micrometers. Current application positioning, three-axis round and elliptical measurement, accessory guides, integration, and Total Vu context.
  3. LaserLinc, Axion One- and Two-Axis Laser Micrometers. Current fit for one- and two-axis, non-round, hot, constrained, and multi-strand applications.
  4. LaserLinc, Sample Inspection. Current Metron and BenchLinc architectures, part-holding options, rotation, full-length inspection, wall measurement, and Total Vu workflows.
  5. LaserLinc, Laser Micrometer Datasheet. V-block, long V-block, feed-through guide, throat-mount, body-mount, plate-mount, rail-mount, adjustable roller, encoder, stand, and accessory information.
  6. LaserLinc, Metron Product Sheet. Published centering cone, magnet V-block, motorized rotation, centering cup, toggle-clamp V-block, cone holder, zero-chuck, throat guide, and motorized part-holder options.
  7. LaserLinc, More Than Measurement. Integrated inline and sample-inspection strategy, automation, Total Vu, adaptable solutions, U.S. manufacturing, and support information.
  8. LaserLinc, Total Vu Process Visualization. Integration of laser, ultrasonic, and FlawSense devices with visualization, analytics, documentation, communication, and control tools.
  9. LaserLinc, Warranty and Service. Typical four-year laser-micrometer and accessory warranty, support terms, software support, and remote Quick Support information.
  10. LaserLinc, About. U.S. ownership, design and manufacturing in Fairborn, Ohio, application engineering, adaptable integrated solutions, and service approach.
  11. NIST, Uncertainty and Dimensional Calibrations. Dimensional uncertainty sources including alignment, Abbe and cosine errors, artifact geometry, support, motion, and thermal effects.
  12. NIST, Dimensional Metrology. Practical fixturing, elastic deformation, temperature, alignment, artifact geometry, and comparator measurement-error context.
  13. NIST, Metrological Traceability. Definition and practical elements of a documented calibration chain in which each calibration contributes to measurement uncertainty.
  14. U.S. FDA, Quality Management System Regulation. QMSR effective February 2, 2026 and incorporation of ISO 13485:2016 by reference.
  15. ISO 10555-1:2023. General requirements for sterile, single-use intravascular catheters. The applicable drawing and product-specific standard govern the dimensional acceptance method.
  16. ISO 11070:2014 and ISO 9626:2016. Guidewire, introducer, dilator, and rigid stainless-steel needle-tubing context for the relevant product families.

Review the part, fixture, and measurement method together

Gauge Advisor is the authorized LaserLinc sales and applications support partner. I help catheter, cannula, guidewire, hypotube, medical tubing, and precision-component teams evaluate, select, quote, integrate, and support LaserLinc micrometers, fixtures, BenchLinc, Metron, UltraGauge, FlawSense, and Total Vu systems.

Gauge Advisor supports equipment selection and integration for the manufacturers represented here; independent process consulting is not offered. Send the part drawing, measurement locations, dimensional range, tolerances, material, length, current setup, photos, data requirements, and representative samples when available. I will respond within one business day, often within a few hours.

  • Micrometer and axis selection
  • V-block, feed-through, roller, and custom fixturing review
  • BenchLinc and Metron architecture comparison
  • Ultrasonic and FlawSense complementary inspection
  • Sample testing and LaserLinc engineering coordination
  • Quotations, integration, training, and ongoing support
Matthew Baker, founder of Gauge Advisor
Founder, Gauge Advisor LLC
Facebook
Twitter
LinkedIn

Need Help Selecting the Right Equipment?

Request Application Guidance

We won’t spam you with emails or call you a thousand times.

Tell us about your process, material, or measurement requirements.

A Gauge Advisor applications specialist will respond within 24 hours to discuss your measurement challenge and help determine the right solution for your line.

No obligation. Just start the conversation.